Cardiac Monitoring with Textile Capacitive Electrodes in Driving Applications: Characterization of Signal Quality and RR Duration Accuracy

  • James Elber Duverger
  • , Geordi Gabriel Renaud Dumoulin
  • , Victor Bellemin
  • , Patricia Forcier
  • , Justine Decaens
  • , Ghyslain Gagnon
  • , Alireza Saidi

Research output: Contribution to journalJournal Articlepeer-review

1 Citation (Scopus)

Abstract

Capacitive ECG sensors in automobiles enable unobtrusive heart rate monitoring as an indicator of a driver’s alertness and health. This paper introduces a capacitive sensor with textile electrodes and provides insights into signal quality and RR duration accuracy. Electrodes of various shapes, sizes, and fabrics were integrated at various positions into the seat back of a driving simulator car seat. Seven subjects completed identical driving circuits with their cardiac signals being recorded simultaneously with textile electrodes and reference Ag-AgCl electrodes. Capacitive ECG signals with observable R peaks (after filtering) could be captured with almost all pairs of textile electrodes, independently of design or placement. Signal quality from textile electrodes was consistently lower compared with reference Ag-AgCl electrodes. Proximity to the heart or even contact with the body seems to be key but not enough to improve signal quality. However, accurate measurement of RR durations was mostly independent of signal quality since 90% of all RR durations measured on capacitive ECG signals had a percentage error below 5% compared to reference ECG signals. Accuracy was actually algorithm-dependent, where a classic Pan–Tompkins-based algorithm was interestingly outperformed by an in-house frequency-domain algorithm.

Original languageEnglish
Article number6097
JournalSensors
Volume25
Issue number19
DOIs
Publication statusPublished - Oct 2025

!!!Keywords

  • ECG sensor
  • ECG signal
  • automobile
  • automotive application
  • capacitive sensing
  • cardiac monitoring
  • driver health status
  • driving application
  • signal quality index
  • textile capacitive electrode

Fingerprint

Dive into the research topics of 'Cardiac Monitoring with Textile Capacitive Electrodes in Driving Applications: Characterization of Signal Quality and RR Duration Accuracy'. These topics are generated from the title and abstract of the publication. Together, they form a unique fingerprint.

Cite this