Skip to main navigation Skip to search Skip to main content

Braided tubular superelastic cables provide improved spinal stability compared to multifilament sublaminar cables

  • École de technologie supérieure
  • Research Center - Hôpital du Sacré-Coeur de Montréal

Research output: Contribution to journal › Journal Article › peer-review

2 Citations (Scopus)

Abstract

This study investigates the use of braided tubular superelastic cables, previously used for sternum closure following sternotomy, as sublaminar fixation method. It compares the biomechanical performance of spinal instrumentation fixation systems with regular sublaminar cables and proprietary superelastic cables. A hybrid experimental protocol was applied to six porcine L1-L4 spinal segments to compare multifilament sublaminar cables (Atlas, Medtronic Sofamor Danek, Memphis, TN) with proprietary superelastic cables. First, intact total range of motion was determined for all specimens using pure moment loading. Second, pure moments were imposed to the instrumented specimens until these intact total ranges of motion were reproduced. Compared to the intact specimens, the use of superelastic cables resulted in stiffer instrumented specimens than the use of multifilament cables for all the loading modes except axial torsion. Consequently, the superelastic cables limited the instrumented segments mobility more than the multifilament cables. Spinal instrumentation fixation systems using superelastic cables could be a good alternative to conventional sublaminar cables as it maintains a constant stabilization of the spine during loading.

Original languageEnglish
Pages (from-to)645-651
Number of pages7
JournalProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
Volume229
Issue number9
DOIs
Publication statusPublished - 5 Sept 2015

!!!Keywords

  • Spine
  • in vitro experiments
  • sublaminar cables
  • superelastic properties

Fingerprint

Dive into the research topics of 'Braided tubular superelastic cables provide improved spinal stability compared to multifilament sublaminar cables'. These topics are generated from the title and abstract of the publication. Together, they form a unique fingerprint.

Cite this