T1 - Calibration, measurement and error analysis of optical temperature measurement via laser induced fluorescence
AU - Harker, Matthew
AU - O'Leary, Paul
AU - Kharicha, Mihaela Stefan
AU - Eck, Sven
PY - 2007
Y1 - 2007
N2 - This paper describes a measurement device for optical temperature measurement. The temperature measurement is based on light induced fluorescence, and hence the device is partly physical and partly theoretical in nature. Rather that estimating a single calibration curve for the measurement, as is the norm with optical measurement, we perform the calibration pixel-wise, which accounts for systematic non-uniformity over the images and measurement cell. With this approach, the measurement is less susceptible to gross outliers. A measurement device can hardly be called such, if the measurement is not accompanied by a measure of its accuracy. We therefore undertake the error propagation throughout the theoretical portion of the measurement device, which leads to a measure of the accuracy of the optical temperature measurement. The analysis not only gives insight into the measurement process, but facilitates the amelioration of the experimental apparatus.
AB - This paper describes a measurement device for optical temperature measurement. The temperature measurement is based on light induced fluorescence, and hence the device is partly physical and partly theoretical in nature. Rather that estimating a single calibration curve for the measurement, as is the norm with optical measurement, we perform the calibration pixel-wise, which accounts for systematic non-uniformity over the images and measurement cell. With this approach, the measurement is less susceptible to gross outliers. A measurement device can hardly be called such, if the measurement is not accompanied by a measure of its accuracy. We therefore undertake the error propagation throughout the theoretical portion of the measurement device, which leads to a measure of the accuracy of the optical temperature measurement. The analysis not only gives insight into the measurement process, but facilitates the amelioration of the experimental apparatus.
KW - Error propagation
KW - Laser induced fluorescence
KW - Mathematical modeling
KW - Measurement accuracy
KW - Optical temperature measurement
UR - https://www.scopus.com/pages/publications/34648824677
U2 - 10.1109/imtc.2007.379008
DO - 10.1109/imtc.2007.379008
M3 - Contribution to conference proceedings
AN - SCOPUS:34648824677
SN - 1424410800
SN - 9781424410804
T3 - Conference Record - IEEE Instrumentation and Measurement Technology Conference
BT - 2007 IEEE Instrumentation and Measurement Technology, IMTC 2007 - Conference Proceedings - Synergy of Science and Technology in Instrumentation and Measurement
PB - Institute of Electrical and Electronics Engineers Inc.
T2 - 2007 IEEE Instrumentation and Measurement Technology, IMTC 2007 - Synergy of Science and Technology in Instrumentation and Measurement
Y2 - 1 May 2007 through 3 May 2007