Skip to main navigation Skip to search Skip to main content

In vivo antegrade fenestration of abdominal aortic stent-grafts

  • Leonard W.H. Tse
  • , Bao T. Bui
  • , Sophie Lerouge
  • , Igor Salazkin
  • , Eric Therasse
  • , Andrew Benko
  • , Hélène Héon
  • , Vincent L. Oliva
  • , Gilles Soulez
  • University of Calgary
  • CIUSSS de l'Estrie-CHUS
  • Centre Hospitalier de L'Universite de Montreal

Research output: Contribution to journal › Journal Article › peer-review

26 Citations (Scopus)

Abstract

Purpose: To examine in a canine model the feasibility of antegrade fenestration of abdominal aortic stent-grafts to preserve the patency of the renal arteries. Methods: Two large dogs underwent antegrade fenestration of stent-grafts in the perirenal aorta. Before fenestration, bare stents were inserted in both renal arteries as fluoroscopic landmarks. A 12-mm iliac extension served as the canine aortic endograft. The first procedure was done under ultrasound and fluoroscopic guidance, using an intravascular ultrasound (IVUS) probe inserted in the vena cava and a Pioneer IVUS catheter. The second was performed exclusively under fluoroscopic guidance with a Brockenbrough needle. Angiograms and duplex ultrasound were planned for 1 month, after which the dogs would be sacrificed for autopsy. The explanted endograft was subjected to biomaterials analysis, with a focus on fabric tear. Results: Perforation of the aortic graft and catheterization of the renal arteries with a floppy guidewire were possible in both animals. In dog 1, aortic graft dilation and subsequent fenestration were not possible, and the experiment was terminated. However, the procedure was successful in both renal arteries of dog 2. At 1-month follow-up in this dog, both renal arteries were patent. Stent fractures were observed bilaterally. There was no extension of the damage to the fabric beyond the area of fenestration. Conclusion: In vivo antegrade fenestration of aortic endografts is technically feasible. However, improvements in technique, instrumentation, and materials are required to make it a reliable and reproducible way of allowing stent-graft vascularization of aortic side branches.

Original languageEnglish
Pages (from-to)158-167
Number of pages10
JournalJournal of Endovascular Therapy
Volume14
Issue number2
DOIs
Publication statusPublished - Apr 2007
Externally publishedYes

!!!Keywords

  • Abdominal aortic aneurysm
  • Angioplasty
  • Aortic aneurysm
  • Canine model
  • Fenestration
  • Intravascular ultrasound
  • Renal artery
  • Stent
  • Stent-graft

Fingerprint

Dive into the research topics of 'In vivo antegrade fenestration of abdominal aortic stent-grafts'. These topics are generated from the title and abstract of the publication. Together, they form a unique fingerprint.

Cite this