Skip to main navigation Skip to search Skip to main content

Safety of plasma-based sterilization: Surface modifications of polymeric medical devices induced by Sterrad® and Plazlyte™ processes

  • S. Lerouge
  • , M. Tabrizian
  • , M. R. Wertheimer
  • , R. Marchand
  • , L. H. Yahia
  • Polytechnique Montréal
  • University of Montreal

Research output: Contribution to journal › Journal Article › peer-review

64 Citations (Scopus)

Abstract

Plasma-based sterilization is a promising alternative to the use of pure ethylene oxide (EO), for low-temperature clinical sterilization of medical instruments and devices. However, few studies have been published that evaluate its safety in terms of possible damage to materials, particularly polymers. The objective of this work was to evaluate polymer surface modifications induced by commercial plasma-based sterilizers, in comparison with pure EO: Samples from 5 polymer-based devices were subjected to 1, 5, and 10 sterilization cycles by Sterrad-100®, Plazlyte™, and pure EO. Surface analysis was carried out by X-ray photoelectron spectroscopy (XPS), dynamic contact angle measurements (DCA), and scanning electron microscopy (SEM). Surface oxidation and wettability changes were observed on all samples sterilized by plasma-based techniques, the degree of modifications depending on the sterilizer (Sterrad®, Plazlyte™ and the type of polymer. Drastic changes of surface appearance were also observed by SEM on PVC samples sterilized by Plazlyte™ and by pure EO. Possible repercussions on safety are discussed.

Original languageEnglish
Pages (from-to)3-13
Number of pages11
JournalBio-Medical Materials and Engineering
Volume12
Issue number1
Publication statusPublished - 2002
Externally publishedYes

Keywords

  • Oxidation
  • Plasma-based sterilization
  • Polymer-based medical device
  • Safety
  • Surface modification

ASJC Scopus subject areas

  • Biomaterials
  • Biomedical Engineering

Fingerprint

Dive into the research topics of 'Safety of plasma-based sterilization: Surface modifications of polymeric medical devices induced by Sterrad® and Plazlyte™ processes'. These topics are generated from the title and abstract of the publication. Together, they form a unique fingerprint.

Cite this