Abstract
Preclinical comparative analysis of biomechanical stability of sternal closure techniques was performed to evaluate the compressive pressure and relative displacement at halves interface in simulated deep breathing or coughing. Artificial sternum models made of polyurethane from Sawbones Corporation, Vashon Island, WA, with a foam density of 0.32 g/cm3, were used. Testing bench force calibration was carried out with a reference load-cell replacing sternum model. The results show that the coughing-mode resembles the deep breathing-mode, with the difference being the speed at which the load is applied. This method permits the acquisition of the compression force applied to the sternum midline and the measurement of its evolution throughout the different testing stages. The gradual decrease in sternum midline compressive forces is explained by the observed local depression at the wire-sternum interface.