Biomechanical assessment of the stabilization capacity of monolithic spinal rods with different flexural stiffness and anchoring arrangement

  • École de technologie supérieure
  • Research Center - Hôpital du Sacré-Coeur de Montréal
  • University of Montreal

Résultats de recherche: Contribution à un journal › Article publié dans une revue, révisé par les pairs › Revue par des pairs

21 Citations (Scopus)

Résumé

Background Spinal disorders can be treated by several means including fusion surgery. Rigid posterior instrumentations are used to obtain the stability needed for fusion. However, the abrupt stiffness variation between the stabilized and intact segments leads to proximal junctional kyphosis. The concept of spinal rods with variable flexural stiffness is proposed to create a more gradual transition at the end of the instrumentation. Method Biomechanical tests were conducted on porcine spine segments (L1-L6) to assess the stabilization capacity of spinal rods with different flexural stiffness. Dual-rod fusion constructs containing three kinds of rods (Ti, Ti-Ni superelastic, and Ti-Ni half stiff-half superelastic) were implanted using two anchor arrangements: pedicle screws at all levels or pedicle screws at all levels except for upper instrumented vertebra in which case pedicle screws were replaced with transverse process hooks. Specimens were loaded in forward flexion, extension, and lateral bending before and after implantation of the fusion constructs. The effects of different rods on specimen stiffness, vertebra mobility, intradiscal pressures, and anchor forces were evaluated. Finding The differences in rod properties had a moderate impact on the biomechanics of the instrumented spine when only pedicle screws were used. However, this effect was amplified when transverse process hooks were used as proximal anchors. Interpretation Combining transverse hooks and softer (Ti-Ni superelastic and Ti-Ni half stiff-half superelastic) rods provided more motion at the upper instrumented level and applied less force on the anchors, potentially improving the load sharing capacity of the instrumentation.

Langue originaleAnglais
Pages (de - à)1026-1035
Nombre de pages10
RevueClinical Biomechanics
Volume30
Numéro de publication10
Les DOIs
ÉtatPublié - 1 déc. 2015

Classification disciplinaire ASJC Scopus

  • Biophysique
  • Orthopédie et médecine sportive

Empreinte digitale

Voici les principaux termes ou expressions associés à « Biomechanical assessment of the stabilization capacity of monolithic spinal rods with different flexural stiffness and anchoring arrangement ». Ces libellés thématiques sont générés à partir du titre et du résumé de la publication. Ensemble, ils forment une empreinte digitale unique.

Citer cette ressource