The aim of the present work was to study corrosion and electrochemical behavior of Ti-22Nb-6Ta (at. %) superelastic alloy in comparison with commercially pure titanium (cp-Ti) under simulated service conditions of load-bearing bone implants, i.e. under cyclic load in simulated body fluid environment. Open circuit potential (OCP) measurements were carried out on wire samples in Hank's solution at 37°C applying bending load (maximum induced strain 0.8%) up to 75×103 cycles or sample failure. It is shown that Ti-22Nb-6Ta alloy is superior to cp-Ti in terms of corrosion fatigue behavior. In particular, it possesses higher OCP values and its passive oxide film is more resistant to the impact of cyclic loading; fatigue life of the alloy is longer.
| Langue originale | Anglais |
|---|---|
| Pages (de - à) | S991-S994 |
| Revue | Materials Today: Proceedings |
| Volume | 2 |
| Numéro de publication | Supplement 3 |
| Les DOIs | |
| État | Publié - 2015 |
Classification disciplinaire ASJC Scopus
- Sciences générales des matériaux
Empreinte digitale
Voici les principaux termes ou expressions associés à « Corrosion fatigue and electrochemical behavior of superelastic Ti-Nb-Ta alloy for medical implants under cyclic load conditions ». Ces libellés thématiques sont générés à partir du titre et du résumé de la publication. Ensemble, ils forment une empreinte digitale unique.Citer cette ressource