This paper focuses on the development of a porous metallic biomimetic femoral stem designed to reduce stress shielding and to provide firm implant fixation through bone ingrowth. The design of this stem starts with the creation of the diagram allowing the establishment of a relationship between the bone ingrowth requirements and the metal additive manufacturing technology limitations. This diagram is then used to determine the optimum porosity (33%) that should compose the porous part of the stem. Afterward, selective laser melting is used to manufacture the porous stem altogether with its fully dense replica. Finite element analysis and numerical homogenization methods are then employed to predict the mechanical behavior of the stem. Both stems are finally tested following the ISO 7206-4 (2010) requirements under static loading conditions. The digital image correlation technique is employed to obtain the displacement and strain fields during the tests, and to validate the finite element model. While the finite element model of the dense stem has been successfully validated, that of the porous stem has shown ~ 40% higher stiffness than that measured experimentally. It has been proven that this discrepancy is due to the difference between the experimentally-measured (42%) and the numerically-targeted (33%) porosities.
| Langue originale | Anglais |
|---|---|
| Pages (de - à) | 546-556 |
| Nombre de pages | 11 |
| Revue | Materials and Design |
| Volume | 114 |
| Les DOIs | |
| État | Publié - 15 janv. 2017 |
SDG des Nations Unies
Ce résultat contribue à ou aux Objectifs de développement durable suivants
-
SDG 7 – Energie propre et d'un coût abordable
-
SDG 9 – Industrie, innovation et infrastructure
-
SDG 12 – Consommation et production durables
Classification disciplinaire ASJC Scopus
- Sciences générales des matériaux
- Mécanique des matériaux
- Ingénierie mécanique
Empreinte digitale
Voici les principaux termes ou expressions associés à « Development of a porous metallic femoral stem: Design, manufacturing, simulation and mechanical testing ». Ces libellés thématiques sont générés à partir du titre et du résumé de la publication. Ensemble, ils forment une empreinte digitale unique.Citer cette ressource
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver