TY - GEN
T1 - Implementation of a 3D porcine lumbar finite element model for the simulation of monolithic spinal rods with variable flexural stiffness
AU - Brummund, Martin
AU - Brailovski, Vladimir
AU - Facchinello, Yann
AU - Petit, Yvan
AU - Mac-Thiong, Jean Marc
N1 - Publisher Copyright:
© 2015 IEEE.
PY - 2015/11/4
Y1 - 2015/11/4
N2 - Monolithic superelastic-elastoplastic spinal rods (MSER) are promising candidates to provide (i) dynamic stabilisation in spinal segments prone to mechanical stress concentration and adjacent segment disease and (ii) to provide fusion-ready stabilization in spinal segments at risk of implant failure. However, the stiffness distributions along the rod's longitudinal axis that best meet clinical requirements remain unknown. The present study is part of a mixed numerical experimental research project and aims at the implementation of a 3D finite element model of the porcine lumbar spine to study the role of MSER material properties and stiffness distributions on the intradiscal pressure distribution in the adjacent segment. In this paper, preliminary intradiscal pressure predictions obtained at one functional spinal unit are presented. Due to a lack of porcine material property data, these predictions were obtained on the basis of uncalibrated human vertebral disc data which were taken from the literature. The results indicate that human annulus and nucleus data predict experimental porcine in vivo and in vitro data reasonably well for the compressive forces of varying magnitudes.
AB - Monolithic superelastic-elastoplastic spinal rods (MSER) are promising candidates to provide (i) dynamic stabilisation in spinal segments prone to mechanical stress concentration and adjacent segment disease and (ii) to provide fusion-ready stabilization in spinal segments at risk of implant failure. However, the stiffness distributions along the rod's longitudinal axis that best meet clinical requirements remain unknown. The present study is part of a mixed numerical experimental research project and aims at the implementation of a 3D finite element model of the porcine lumbar spine to study the role of MSER material properties and stiffness distributions on the intradiscal pressure distribution in the adjacent segment. In this paper, preliminary intradiscal pressure predictions obtained at one functional spinal unit are presented. Due to a lack of porcine material property data, these predictions were obtained on the basis of uncalibrated human vertebral disc data which were taken from the literature. The results indicate that human annulus and nucleus data predict experimental porcine in vivo and in vitro data reasonably well for the compressive forces of varying magnitudes.
UR - https://www.scopus.com/pages/publications/84953216968
U2 - 10.1109/EMBC.2015.7318512
DO - 10.1109/EMBC.2015.7318512
M3 - Contribution to conference proceedings
C2 - 26736412
AN - SCOPUS:84953216968
T3 - Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS
SP - 917
EP - 920
BT - 2015 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2015
PB - Institute of Electrical and Electronics Engineers Inc.
T2 - 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2015
Y2 - 25 August 2015 through 29 August 2015
ER -