LPBF-fabricated superelastic Ti–Zr–Nb open cell lattice structures for orthopedic implants: Process optimization, geometric conformity, microstructure, static and fatigue mechanical behavior

  • V. Sheremetyev
  • , V. Lezin
  • , K. Lukashevich
  • , S. Chernyshikhin
  • , V. Cheverikin
  • , A. Tvorogov
  • , M. Derkach
  • , S. Prokoshkin
  • , V. Brailovski
  • National University of Science and Technology "MISiS"
  • Lomonosov Moscow State University

Résultats de recherche: Contribution à un journal › Article publié dans une revue, révisé par les pairs › Revue par des pairs

1 Citation (Scopus)

Résumé

Additively manufactured orthopedic implants are required to simultaneously offer a high geometric accuracy, a low elastic modulus, and an excellent fatigue resistance. In this study, we optimized laser powder bed fusion (LPBF) process parameters to fabricate three open cell lattice structures from a superelastic Ti–Zr–Nb alloy: a rhombic dodecahedron, a sheet gyroid, and a Voronoi polyhedral. In these structures, the strut thickness ranged between 200 and 400 μm and the functional porosity, between 65 and 80%. Geometric conformity and defect porosity were quantified using microcomputed tomography, revealing that the optimized processing limited the maximum deviations in functional porosity to 9%, with the gyroid architecture demonstrating the highest dimensional accuracy. At the same time, the defect porosity in all the structures did not exceed 0.2%. After the heat treatment, microstructural analysis and mechanical testing confirmed the presence in these structures of metastable β-phase. Under quasi-static compression, the lattice specimens manifested a yield plateau associated with stress-induced martensitic transformation. Fatigue testing of Ti–Zr–Nb structures demonstrated a stable cyclic performance and competitive fatigue resistance up to 5 × 106 cycles as compared to their benchmark Ti–6Al–4V counterparts. The results obtained highlight the potential of LPBF-fabricated superelastic Ti–Zr–Nb lattice specimens for load-bearing implant applications, owing to their reduced stiffness and enhanced long-term mechanical durability.

Langue originaleAnglais
Pages (de - à)4014-4032
Nombre de pages19
RevueJournal of Materials Research and Technology
Volume43
Les DOIs
ÉtatPublié - 1 juil. 2026

SDG des Nations Unies

Ce résultat contribue à ou aux Objectifs de développement durable suivants

  1. SDG 9 – Industrie, innovation et infrastructure
    SDG 9 – Industrie, innovation et infrastructure

Classification disciplinaire ASJC Scopus

  • Céramiques et composites
  • Biomatériaux
  • Surfaces, revêtements et films
  • Métaux et alliages

Empreinte digitale

Voici les principaux termes ou expressions associés à « LPBF-fabricated superelastic Ti–Zr–Nb open cell lattice structures for orthopedic implants: Process optimization, geometric conformity, microstructure, static and fatigue mechanical behavior ». Ces libellés thématiques sont générés à partir du titre et du résumé de la publication. Ensemble, ils forment une empreinte digitale unique.