Skip to main navigation Skip to search Skip to main content

Improving greater trochanteric reattachment with a novel cable plate system

  • École de technologie supérieure
  • Research Center - Hôpital du Sacré-Coeur de Montréal
  • University of Alberta

Research output: Contribution to journal › Journal Article › peer-review

9 Citations (Scopus)

Abstract

Cable-grip systems are commonly used for greater trochanteric reattachment because they have provided the best fixation performance to date, even though they have a rather high complication rate. A novel reattachment system is proposed with the aim of improving fixation stability. It consists of a Y-shaped fixation plate combined with locking screws and superelastic cables to reduce cable loosening and limit greater trochanter movement.The novel system is compared with a commercially available reattachment system in terms of greater trochanter movement and cable tensions under different greater trochanteric abductor application angles.A factorial design of experiments was used including four independent variables: plate system, cable type, abductor application angle, and femur model. The test procedure included 50 cycles of simultaneous application of an abductor force on the greater trochanter and a hip force on the femoral head.The novel plate reduces the movements of a greater trochanter fragment within a single loading cycle up to 26%. Permanent degradation of the fixation (accumulated movement based on 50-cycle testing) is reduced up to 46%. The use of superelastic cables reduces tension loosening up to 24%. However this last improvement did not result in a significant reduction of the grater trochanter movement.The novel plate and cables present advantages over the commercially available greater trochanter reattachment system. The plate reduces movements generated by the hip abductor. The superelastic cables reduce cable loosening during cycling. Both of these positive effects could decrease the risks related to grater trochanter non-union.

Original languageEnglish
Pages (from-to)383-391
Number of pages9
JournalMedical Engineering and Physics
Volume35
Issue number3
DOIs
Publication statusPublished - 2013

Keywords

  • Cable tension
  • Greater trochanter reattachment
  • Implant
  • Locking screw
  • Orthopaedic
  • Slide osteotomy

ASJC Scopus subject areas

  • Biophysics
  • Biomedical Engineering

Fingerprint

Dive into the research topics of 'Improving greater trochanteric reattachment with a novel cable plate system'. These topics are generated from the title and abstract of the publication. Together, they form a unique fingerprint.

Cite this