Abstract
Cerclage cables, frequently used in the management of fractures and osteotomies, are associated with a high failure rate and significant loosening during surgery. This study compared the capacity to maintain tension of different types of orthopaedic cable systems. Multifilament Cobalt-Chrome (CoCr) cables with four different crimp/clamp devices (DePuy, Stryker, Zimmer and Smith&Nephew) and one non-metallic Nylon (Ny) cable from Kinamed were instrumented with a load cell to measure tension during insertion. Significant tension loss was observed with crimping for all cables (P<. 0.05). Removing the tensioner led to an additional unexpected tension loss (CoCr-DePuy: 18%, CoCr-Stryker: 29%, CoCr-Smith&Nephew: 33%, Ny: 46%, and CoCr-Zimmer: 52%). The simple CoCr (DePuy) cable system outperformed the more sophisticated locking devices due to its significantly better ability to prevent tension loss.
| Original language | English |
|---|---|
| Pages (from-to) | 1509-1512 |
| Number of pages | 4 |
| Journal | Journal of Arthroplasty |
| Volume | 28 |
| Issue number | 9 |
| DOIs | |
| Publication status | Published - Oct 2013 |
Keywords
- Cable
- Cerclage
- Fractures
- Osteotomie
- Tension loss
ASJC Scopus subject areas
- Orthopedics and Sports Medicine
Fingerprint
Dive into the research topics of 'Initial tension loss in cerclage cables'. These topics are generated from the title and abstract of the publication. Together, they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver