Initial tension loss in cerclage cables

  • Research Center - Hôpital du Sacré-Coeur de Montréal
  • École de technologie supérieure

Résultats de recherche: Contribution à un journal › Article publié dans une revue, révisé par les pairs › Revue par des pairs

12 Citations (Scopus)

Résumé

Cerclage cables, frequently used in the management of fractures and osteotomies, are associated with a high failure rate and significant loosening during surgery. This study compared the capacity to maintain tension of different types of orthopaedic cable systems. Multifilament Cobalt-Chrome (CoCr) cables with four different crimp/clamp devices (DePuy, Stryker, Zimmer and Smith&Nephew) and one non-metallic Nylon (Ny) cable from Kinamed were instrumented with a load cell to measure tension during insertion. Significant tension loss was observed with crimping for all cables (P<. 0.05). Removing the tensioner led to an additional unexpected tension loss (CoCr-DePuy: 18%, CoCr-Stryker: 29%, CoCr-Smith&Nephew: 33%, Ny: 46%, and CoCr-Zimmer: 52%). The simple CoCr (DePuy) cable system outperformed the more sophisticated locking devices due to its significantly better ability to prevent tension loss.

Langue originaleAnglais
Pages (de - à)1509-1512
Nombre de pages4
RevueJournal of Arthroplasty
Volume28
Numéro de publication9
Les DOIs
ÉtatPublié - oct. 2013

Classification disciplinaire ASJC Scopus

  • Orthopédie et médecine sportive

Empreinte digitale

Voici les principaux termes ou expressions associés à « Initial tension loss in cerclage cables ». Ces libellés thématiques sont générés à partir du titre et du résumé de la publication. Ensemble, ils forment une empreinte digitale unique.

Citer cette ressource